1–5 Sept 2008
University of Glasgow
Europe/London timezone

Image-guided radiotherapy using active pixel technology

2 Sept 2008, 14:50
20m
University of Glasgow

University of Glasgow

Glasgow G12 8QQ UK
Oral Contribution Applications in Nuclear Medicine and Radiology Applications in Nuclear Medicine and Radiology

Speaker

John Osmond

Description

The intelligence, read-out speed, radiation hardness and large size of CMOS active pixel sensors (APS) gives them a potential advantage over current radiotherapy verification systems. This work investigates the feasibility of using an APS to image the megavoltage treatment beam produced by a linear accelerator, and demonstrates the logic which may be used to evaluate treatment and track motion. A CMOS APS was incorporated into an imaging system, and anatomical imaging, resolution and contrast were evaluated. Two algorithms were used to determine the field-area, delivered dose and the position of collimator leaves in an intensity-modulated radiotherapy treatment. Results agreed with the prescription to within a single frame for dose delivery, and 0.03 cm for the position of collimator leaves. Such a system therefore shows potential for online verification.

Author

Presentation materials