(G) (POS-28) A wireless resonator layer for high-resolution TMJ MRI at 1.5 T

28 May 2024, 18:07
2m
PAB Hallways (Western University)

PAB Hallways

Western University

Poster not-in-competition (Graduate Student) / Affiche non-compétitive (Étudiant(e) du 2e ou 3e cycle) Physics in Medicine and Biology / Physique en médecine et en biologie (DPMB-DPMB) DPMB Poster Session & Student Poster Competition (28) | Session d'affiches DPMB et concours d'affiches étudiantes (28)

Speakers

Mr Bocheng Xu Yujie Ren (University of Winnipeg)

Description

INTRODUCTION
MRI provides highly detailed images that enable healthcare professionals to assess the joints and surroundings in great detail. While commercial MRI scanners typically come equipped with basic receive coils, such as the head receive array, RF coils tailored for specialized applications like TMJ MRI must be obtained separately. Consequently, TMJ MRI scans often use suboptimal head receive array 1-4 due to the lack of specialized coils.

In this study, we introduce a simple, low-cost, and easy-to-reproduce wireless resonator insert to enhance the quality of TMJ MRI at 1.5 Tesla. The wireless resonator shows a significant improvement in SNR and noticeably better imaging quality compared to the head array alone in both phantom and in vivo images.

METHODS
Figures 1A and 1B show the head neck receive array and a wireless resonator. Figure 1C illustrates the position of wireless resonator for TMJ MRI. Figure 1D depicts the circuit diagram of the wireless resonator which was tuned to 63.67 MHz, the passive detune circuit disables the wireless resonator during the transmit phase, similar to the designs in previous works 5-8 The centers of the wireless resonator pads are aligned with TMJ for optimal imaging performance. The body coil is used for RF transmission, while the head array is employed for RF reception.
We perform multiple tests to assess the performance with and without the wireless resonator inserted into the head array:
• The transmit field (B1+) map and RF power calibration for detuning effectiveness
• Phantom image for SNR measurement
• Volunteer image for clinical evaluation
The wireless resonator operates in receive-only mode, modifying the scanners’ default parameter settings is unnecessary
Local board approved human procedures; participant provided written consent. Safety test conducted before imaging9.
RESULTS
Figure 2 compares axial B1+ with and without the wireless resonator insert. The difference between these two B1+ maps is <1%. Additionally, The RF power change for a 180-degree flip angle was under 1.5% with and without the wireless resonator. These affirm that the wireless insert remains highly transparent to RF power during the transmit phase.
In the context of TMJ MRI, where we typically focus on anatomical structures like the articular fossa, articular eminence, and disc, the average depth rarely exceeds 2.5 cm10. The SNR improvement (averaged over the red box in Figures 3C and 3D) achieved with the wireless resonator can reach up to 5.3 times at this depth. the SNR (averaged over the yellow box in Figures 3C and 3D), remains 2.4 times even at a depth of 4 cm.
Figure 4 displays volunteer TMJ images, acquired using multi-slice sagittal T1-weighted and PWD images. Combining the wireless resonator with the head array significantly improves image quality over using the head array alone. This aligns with our phantom study, where the wireless resonator consistently provided higher SNR. To achieve acceptable quality with just the head array, thicker slices or longer scan times are necessary11.

DISCUSSIONS
We chose the head array instead of the body coil as the primary coil for following reasons:
• It offers stronger mutual coupling, higher wireless power transfer efficiency, lower coil loss.
• This choice combines large array coverage with high local SNR, aiding TMJ MRI localization.
• Parallel imaging functionality. In Figure 5, the g-factor computations performed on a phantom with R-L acceleration factors ranging from 2 to 4. The comparison with g-factors obtained from a head array is also provided. The majority of commercially available TMJ coils do not possess parallel imaging capabilities11,12.

Prioritizing safety and comfort, a flexible printed circuit board coil is securely embedded in an MRI-compatible foam pad, which is only 20 mm thick and adaptable to different anatomical shapes. The coil operates exclusively in receive-only mode, avoiding interference with transmit signals and eliminating hotspots. For added safety, a fuse is included in the passive detune circuit in case of malfunction.

CONCLUSION
The combination of wireless RF resonators and phased arrays enhances SNR in specific regions and enables parallel imaging within existing MRI setups.

This approach could prove beneficial for imaging other anatomies, such as the thyroid, eye, and carotid artery. Different wireless RF resonators can also be integrated with diverse receive arrays to acquire extremity, breast, and body images tailored to specific anatomies.

Beyond using L/C resonators for wireless inserts, alternative solutions may involve volume-type wireless resonators or metamaterial-inspired designs 13,14

This advancement ensures affordability, streamlined workflow, and flexibility across different magnetic field strengths.

REFERENCES

  1. D.T. dos Santos, M.G.P. Cavalcanti, Osteosarcoma of the temporomandibular joint: Report of 2 cases, Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology. 94 (2002) 641–647. https://doi.org/10.1067/moe.2001.129181.
  2. R. Emshoff, K. Innerhofer, A. Rudisch, S. Bertram, Clinical versus magnetic resonance imaging findings with internal derangement of the temporomandibular joint: An evaluation of anterior disc displacement without reduction, Journal of Oral and Maxillofacial Surgery. 60 (2002) 36–41. https://doi.org/10.1053/joms.2002.29071.
  3. Y. Uchiyama, K. Matsumoto, S. Murakami, T. Kanesaki, A. Matsumoto, M. Kishino, S. Furukawa, MRI in a case of osteosarcoma in the temporomandibular joint, Dentomaxillofacial Radiology. 43 (2014) 20130280. https://doi.org/10.1259/dmfr.20130280.
  4. S. Krohn, N. Gersdorff, T. Wassmann, K.-D. Merboldt, A.A. Joseph, R. Buergers, J. Frahm, Real-time MRI of the temporomandibular joint at 15 frames per second—A feasibility study, European Journal of Radiology. 85 (2016) 2225–2230. https://doi.org/10.1016/j.ejrad.2016.10.020.
  5. H. Zhu, M. Fallah-Rad, M. Lang, W. Schellekens, K. Champagne, L. Petropoulos, A novel multichannel wireless receive phased array coil without integrated preamplifiers for high field MR imaging applications, in: ISMRM, 2012: p. 2788.
  6. H. Zhu, M. Fallah-Rad, L. Petropoulos, Wireless phased array endorectal coil for prostate imaging, in: 2013: p. 2732.
  7. M. Lu, S. Chai, H. Zhu, X. Yan, Low-cost inductively coupled stacked wireless RF coil for MRI at 3 T, NMR in Biomedicine. 36 (2023) e4818. https://doi.org/10.1002/nbm.4818.
  8. IEC 60601-1:2023 SER | IEC Webstore | electromagnetic compatibility, EMC, smart city, (n.d.). https://webstore.iec.ch/publication/2603 (accessed July 27, 2023).
  9. A. Manoliu, G. Spinner, M. Wyss, S. Erni, D.A. Ettlin, D. Nanz, E.J. Ulbrich, L.M. Gallo, G. Andreisek, Quantitative and qualitative comparison of MR imaging of the temporomandibular joint at 1.5 and 3.0 T using an optimized high-resolution protocol, Dentomaxillofacial Radiology. 45 (2016) 20150240. https://doi.org/10.1259/dmfr.20150240.
  10. E.J. Inarejos Clemente, M. Tolend, T. Junhasavasdikul, J. Stimec, N. Tzaribachev, B. Koos, L. Spiegel, R. Moineddin, A.S. Doria, Qualitative and semi-quantitative assessment of temporomandibular joint MRI protocols for juvenile idiopathic arthritis at 1.5 and 3.0T, European Journal of Radiology. 98 (2018) 90–99. https://doi.org/10.1016/j.ejrad.2017.10.024.
  11. R. Schmidt, A. Slobozhanyuk, P. Belov, A. Webb, Flexible and compact hybrid metasurfaces for enhanced ultra high field in vivo magnetic resonance imaging, Sci Rep. 7 (2017) 1678. https://doi.org/10.1038/s41598-017-01932-9.
  12. V. Puchnin, G. Solomakha, A. Nikulin, A.W. Magill, A. Andreychenko, A. Shchelokova, Metamaterial inspired wireless coil for clinical breast imaging, Journal of Magnetic Resonance. 322 (2021) 106877. https://doi.org/10.1016/j.jmr.2020.106877.
Keyword-1 TMJ
Keyword-2 MRI
Keyword-3 wireless coil