The primary objective was to implement VMAT planning technique for tbi patients and evaluate its feasibility across three key metrics: PTV coverage, the Homogeneity Index, and OAR sparing, while validating our delivery accuracy through EPID-based dosimetric verification.
Result
The results were highly encouraging. For all five patients, 90%of the prescribed dose was received by an average of over 97% of volume while keeping the hot spots under control, and controlled OAR doses. Furthermore, our calculated Homogeneity Index remained stable across the cohort, averaging approximately 0.21.
Background
Conventionally, TBI is delivered using the extended SSD technique, using large AP–PA or lateral fields. With recent advancements in radiotherapy, IMRT, VMAT, and helical tomotherapy have emerged as alternatives for TBI delivery for improved target coverage, reproducible delivery, and most importantly effectively spare critical organs.
Discussion (methods + interpretation)
Treatment planning was done on the CT data sets of five patients using 5-6 isocenters on Eclipse 15.6. Whole body was considered as PTV excluding lens, lungs and kidneys. A 6MV photon beam was used for all the plans. The planning goal was for 95%-99% of the PTV to receive at least 90% of the prescribed dose minimizing dose to OARs and controlling hotspots. Patient-specific quality assurance (QA) was performed using EPID portal dosimetry to verify the accuracy of the delivered treatment plans.