Comparative Analysis of Beam Profiles and Dosimetric Characteristics of a Standard Clinical Linear Accelerator and the HEELA-Developed Prototype Linac

21 Nov 2026, 14:00
5m
Online Conference

Online Conference

Online Conference

Speaker

Mr DANYAL AHMAD (Pakistan institute of engineering and applied sciences PIEAS)

Aim & Purpose

This study aimed to compare the dosimetric characteristics and beam profiles of the HEELA-developed prototype linac and the Varian IX clinical linac at IRNUM, Peshawar Pakistan. The objective was to establish a dosimetric baseline, assess beam quality, flatness and symmetry, and identify areas requiring optimisation.

Result

The Varian IX linac exhibited a TPR₂₀,₁₀ of 0.6645, compared with 0.6513 for the HEELA prototype, indicating a difference in beam quality. The clinical linac met the applied flatness and symmetry tolerances, whereas HEELA exceeded them at all measured depths. HEELA flatness ranged from 3.98% to 6.96%, and symmetry ranged from 2.08% to 3.80%. Despite these profile non-uniformities, the prototype demonstrated stable output and acceptable absolute calibration, providing a baseline for further engineering optimisation.

Discussion (methods + interpretation)

Both linacs were evaluated using nominal 6 MV photon beams under with-flattening-filter conditions. Absolute output calibration and beam quality measurements were performed according to the IAEA TRS-398 protocol using a PTW Farmer-type ionisation chamber. An Exradin A28 chamber in a three-dimensional water phantom was used to acquire in-plane and cross-plane beam profiles at multiple depths. Percentage depth dose, TPR₂₀,₁₀, flatness and symmetry were assessed, with beam flatness and symmetry evaluated against AAPM TG-142 tolerance criteria. The measurements enabled comparison of beam quality and profile uniformity between the two systems.

Background

Accurate beam dosimetry is essential for safe and effective radiotherapy. Medical linear accelerators must deliver consistent radiation doses with acceptable beam quality and spatial uniformity. The HEELA-developed prototype linac represents a national effort to develop indigenous radiotherapy technology in Pakistan. However, its dosimetric characteristics require systematic comparison with an established clinical linear accelerator.