Speaker
Description
The application of nonlinear calibration techniques is popular in the evaluation of the response exhibited by chemical and biosensors, and yet the dependence of resultant parameters on the chosen concentration range for calibration is frequently ignored. Herein, we conduct a computational study of the data already reported in the literature regarding the calibration of an organic thin-film transistor with a Ti3C2Tx MXene gate in the detection of dopamine. In this study, the data already published is subjected to an independent computer analysis based on a Python program with various calibration concentration windows. Various constants, such as sensitivity and response constants dependent on the calibration, are examined in detail, with the measured values of the given parameters being compared with each other. Further on, the effect of the chosen response model on the calibration parameters is analyzed. The results indicate that the calibration parameters are strongly affected both by the calibration concentration window and the calibration model, especially when the gathered data demonstrate different concentration-dependent behaviors. The obtained conclusions suggest that the calibration parameters should be analyzed considering the concentration window and used response model rather than as some objective parameters of the sensor. The study provides a reproducible computational framework for assessing the robustness of calibration-derived parameters and for improving the quantitative interpretation and comparison of MXene-based biochemical sensor responses.