Mixed helium–carbon beams enable online range monitoring because both ions share the same mass-to-charge ratio and can be accelerated simultaneously, while helium has approximately three times the range in water. Anatomical changes beyond the carbon Bragg peak therefore affect helium, but not carbon. We developed a deep learning approach for spot-wise carbon range prediction using mixed-beam detector signals and planning CT information.
Methods
Using our Geant4-based GPU Monte Carlo code Raymond, 882,350 mixed helium–carbon pencil beams with 10% helium were simulated for two-field plans of 17 lung and liver patients in end-inhale and end-exhale phases. Detector images were first processed by a defragmentation network. A second network combined cleaned detector images, beam energy, planning CT and reference carbon dose profile to predict the carbon dose profile in the second respiratory phase and derive the R80 range change.
Results
Five-fold cross-validation yielded MAE/RMSE values of 2.0/4.0 mm for lung and 1.3/2.2 mm for liver. Assuming no carbon range change resulted in 3.7/6.3 mm and 2.2/3.7 mm, while assuming equal helium and carbon range changes gave 4.6/8.2 mm and 2.1/3.2 mm, respectively. This corresponds to MAE reductions of 48–58% for lung and 38–43% for liver. Total inference time was ~7 ms per pencil beam on an NVIDIA H200.
Outlook & Discussion
These results demonstrate the potential of mixed beams for real-time, spot-by-spot carbon range monitoring. Experimental validation and evaluation of clinically relevant dose changes are planned.