Mixed beams hold a great potential for treatment feedback in ion beam therapy. One central question besides the generation of mixed beams resides in what kind of detection system should be used. Ideally, the system should readily keep up with the clinical beam intensities and spot-switching times, be sensitive to even small content of the imaging ion in the beam, and importantly, generate actionable, real-time feedback on the treatment quality. This presentation highlights some clinical constraints, before a comparison of two commonly used system design choices in particle imaging is presented: single-event and integration mode imaging. While the former provides very good image resolution, the latter is comparatively more affordable an less complex. In addition, current single-event systems are not ready to face the expected high rate of fragments produced by the treatment beam. Current efforts with integrating mode systems in mixed beams at GSI and MedAustron are presented. Finally, an outlook is provided for future systems.