Our work is part of the effort toward the direct detection of dark matter within the Recoil Directionality (ReD) experiment, conducted by a subgroup of the DarkSide Collaboration. For this purpose, a Neutron Generator (NG), based on deuterium-deuterium (D-D) fusion, was commissioned at the Institute of Physics, University of São Paulo. This generator will be used as a neutron source for nuclear recoil measurements in a Liquid Argon Time Projection Chamber (LArTPC).
During the latest data acquisitions with the NG, a degradation in the energy resolution of the silicon detector (SiDet) - which detects ${}^3\mathrm{He}$, protons, and tritium produced in D-D fusion - was observed, along with a progressive shift of the energy peaks of these components toward lower energies over time. These effects may be associated with detector damage caused by elastically scattered deuterium ions, suggesting that the protective film then in use, consisting of an aluminium layer deposited on a carbon backing, did not provide sufficient shielding. The objective of this work is to verify this hypothesis and determine an improved protective-film configuration.
To this end, SRIM simulations were performed in three stages: evaluation of deuterium incidence on the SiDet without protection, analysis of the performance of the film previously in use, and determination of the ideal thicknesses for new films. The simulations showed that, in the absence of protective films, $99.93\%$ of the incident deuterium ions reached the active region of the detector and confirmed that the previous film was ineffective in preventing deuterium ions from reaching this region. Since thin metallic films may contain microscopic imperfections (holes), a configuration using two aluminium films, each supported by a carbon backing, was investigated. The thicknesses of both films were then determined through simulation with the goal of reducing deuterium transmission below $0.1\%$.
Based on these results, the new films were manufactured and tested during NG operation. New data show that the ${}^3\mathrm{He}$, proton, and tritium energy peaks are no longer shifting over time and the resolution is no longer degrading, confirming that the new protective film configuration effectively prevents damage to the silicon detector during system operation.
| Keywords | Dark matter, Neutron Generator, Silicon detector, SRIM simulations |
|---|---|
| Proceedings Publication Intent | No |
| Student Award Eligibility | yes |
| Financial Support Request | no |