Experimental Characterization of a $2^{\prime\prime} \times 2^{\prime\prime}$ BGO Scintillation Detector with Integrated SiPM

Dec 9, 2026, 11:00 AM
1h 40m
IPEN

IPEN

Cidade Universitária - Bairro Butantã CEP: 05508-000 - São Paulo - SP
Poster Nuclear Instrumentation Poster

Speakers

Julian Estupiñan (Student) Samuel Posada

Description

Inorganic scintillators such as BGO are widely used in gamma-ray spectroscopy, medical physics, and high-energy physics due to their high density and high effective atomic number. The experimental characterization of these detectors is essential to validate their performance and establish their range of applicability.

In this work, the experimental characterization of a $2^{\prime\prime} \times 2^{\prime\prime}$ BGO (Bismuth Germanate) scintillator with an integrated silicon photomultiplier (SiPM) was carried out. Amptek's DP5 acquisition electronics unit and its accompanying DppMCA software were employed.

The procedure consisted of analyzing the spectra of different radiation sources, placed \SI{20.1(1)}{\centi\meter} from the detector, using the information obtained from Gaussian fits performed on the characteristic photopeak of each source. This procedure covered an energy range from \SI{81}{\keV} ($^{133}\text{Ba}$) to \SI{1836}{\keV} ($^{88}\text{Y}$). For this purpose, an energy calibration was carried out, i.e., the channel-to-energy conversion was fitted.

The results of this analysis allowed the determination of the detector's linearity. Additionally, fits were performed for both the energy resolution and the efficiency, allowing their behavior to be described as a function of energy, with good agreement observed in both cases. Furthermore, an experimental method was used to determine the detector's response function, allowing its behavior to be known for any energy value within the studied range.

This work presents a complete characterization of the $2^{\prime\prime} \times 2^{\prime\prime}$ BGO detector, providing a solid basis for its application in nuclear spectroscopy, environmental radiation measurement, and future characterization of similar detectors.

Keywords BGO, SiPM, detectors, scintillators, resolution, efficiency, gamma-ray spectroscopy