DAILY DOSE TEST RESULT EVALUATION OF THE PRECISE ELEKTA ACCELERATORS AT HO CHI MINH CITY ONCOLOGY HOSPITAL

8 Sept 2022, 11:30
15m

Speaker

Nguyễn Khắc Sơn .

Description

Target: Evaluation of daily dose test results of the Precise Elekta accelerators, there
by make conclusions about the dose stability of Precise Elekta accelerators.
Subjects, methods: Perform dose measurement on Precise Elekta accelerators
using Daily QA3 device based TRS-398[1], TG-142[2] reports and reference guide
of Sun Nuclear Corporation.
Results: In this study, results of daily dose measurement were obtained for 331 days
(18 months), from January 2, 2018 to June 30, 2019. In general, the energy levels
of the Presice Elekta accelerators bring good dose quality, the number of days
exceeding the limit is very low. Common energy levels such as 6MV and 9MeV are
always highly stable, highly accurate, and will never exceed the limit.
Conclusions: The Presice Eleckta accelerators at Ho Chi Minh City Oncology
Hospital is highly appreciated in terms of dose, the dose measuring device works
well, is operated by a team of highly qualified engineers, always bringing the best
results for radiationtherapy patients.