Accurate energy reconstruction in scintillator detectors requires a consistent treatment of channel-to-channel response variations, multiple preamplifier gains, and the non-uniform sharing of scintillation light across a photosensor array. These effects are particularly important for monolithic scintillators read out by multiple SiPM channels, where the reconstructed energy is obtained from their combined response.
We present a four-stage calibration strategy for a monolithic 3 mm thick GAGG:Ce scintillator coupled to a 4 × 4 SiPM array and read out using the dual-gain CITIROC ASIC based CAEN electronics. The procedure begins with pedestal subtraction, single-photoelectron equalisation, and zero-suppression derived from dedicated data collected in the absence of any source. A channel-by-channel high-gain/low-gain harmonisation then places both gain paths on a common high-gain-equivalent scale, allowing measurements beyond high-gain saturation to be recovered from the low-gain response. Residual differences in channel response arising from gain and light-collection variations are subsequently corrected using a regularised minimum-variance inspired alignment based on the 511 keV Na-22 photopeak. Finally, the summed response is converted to energy using a quadratic ADC-to-MeV calibration derived from six Eu-152 photopeaks spanning 122–1408 keV.
The calibrated response is evaluated using independent Na-22 and Co-60 measurements. The resulting energy scales are calibrated with a linearity better than 1%. The minimum variance algorithm based alignment also improves the relative energy resolution of the 511 keV peak width from 11.62% to 11.01%. This study thus demonstrates a practical staged approach for calibrating multi-channel SiPM-based scintillator detectors over a broad energy range while explicitly accounting for dual-gain readout and spatially non-uniform light sharing.
| Tracks | Accelerator Physics, Instrumentation and Future Experiments |
|---|