Instrumentation and commissioning of the expanded GRAPES-3 muon telescope with recycled legacy proportional tubes

Not scheduled
20m
CCE (IIT Hyderabad)

CCE

IIT Hyderabad

https://physics.iith.ac.in/daehep2026/
Poster

Speaker

Dr Pranaba K Nayak (Tata Institute of Fundamental Research)

Description

The GRAPES-3 experiment is nearly doubling its muon telescope area from 560 m² to ∼1130 m² by refurbishing ∼3,300 legacy proportional counters from the Kolar Gold Field experiment [1]. This work presents the instrumentation methodology developed for this expansion, built on three pillars: material characterization, high-throughput vacuum processing, and long-term operational feedback.

First, non-destructive characterization using XRD, FTIR, and SEM-EDX assessed the surface condition of half-century-old iron tubes, validated cleaning protocols, and established acceptance criteria for detector refurbishment and qualification [1–3]. Second, vacuum manifolds evolved through four generations (Gen-I–IV) for parallel detector processing. The optimized SS316 system achieved 3.6×10⁻⁷ mbar and enabled evacuation, leak testing, degassing, and P10 gas filling of ∼3,300 counters at ∼10/day [4], with >97% satisfying acceptance criteria. Third, weekly remote performance meetings over 12 years (2011–2022) provided operational feedback from the existing telescope, directly informing refurbishment procedures, verification methods, acceptance criteria, and integration strategy.

Performance verification using spectroscopic amplifiers and a MAESTRO-based multi-channel analyzer (MCA) demonstrated stable detector response; six-month monitoring showed count-rate variations below 2%, while new front-end electronics reduced the noise floor by up to 30% [5–7]. The expanded system is not yet fully commissioned. Partial overburden currently provides a vertical muon threshold of ∼0.5 GeV, expected to increase to ∼1 GeV after completion [5,6]. Full validation and physics measurements will follow final commissioning. The demonstrated methodology provides a scalable route for recovering, qualifying, and integrating large populations of legacy gaseous detectors for future HEP experiments.
References:
[1] P.K. Nayak et al., Proc. XXVI DAE-BRNS High Energy Physics Symposium 2024, Springer Proc. Phys., vol. 432 (2025) 748–751. (Poster award)
[2] P.K. Nayak et al., Next-Generation Materials and Devices: Select Proceedings of ICNMD 2024, Chap. 8, Springer Proc. Mater. (2026).
[3] P.K. Nayak et al., Zast. Mater. 66 (2025) 887.
[4] P.K. Nayak et al., Nucl. Instrum. Methods Phys. Res. A 1093 (2027) 171971.
[5] P.K. Mohanty, Eur. Phys. J. Spec. Top. 234 (2025) 5009.
[6] P.K. Mohanty, Nucl. Instrum. Methods Phys. Res. A 1093 (2027) 171992.
[7] K. Ramesh, P.A. Wincy, Phys. At. Nucl. 88 (2025) S274.

Tracks Accelerator Physics, Instrumentation and Future Experiments

Author

Dr Pranaba K Nayak (Tata Institute of Fundamental Research)

Co-authors

Dr PK Mohanty (Tata Institute of Fundamental Research) Dr SK Gupta (Tata Institute of Fundamental Research) Dr Y Hayashi (Graduate School of Science, Osaka Metropolitan University, Osaka, 558 8585, Japan) Dr S Kawakami (Graduate School of Science, Osaka Metropolitan University, Osaka, 558 8585, Japan) Dr SR Dugad (Tata Institute of Fundamental Research) Mr A Jain (Tata Institute of Fundamental Research) Mr P Jagadeesan (Tata Institute of Fundamental Research) Mr K Ramesh (Tata Institute of Fundamental Research) Dr M Muthuvinayagam (Department of Physics, Vel Tech Rangarajan Dr Sagunthala R&D Inst. of Sci & Tech., Chennai 600 062, India) Dr A Oshima (College of Engineering, Chubu University, Kasugai, 487 8501, Japan) Mr SD Morris (Tata Institute of Fundamental Research)